TY - JOUR
T1 - Temperature distributions during clinical scanned, focused ultrasound hyperthermia treatments
AU - Hynynen, K.
AU - Shimm, D.
AU - Anhalt, D.
AU - Stea, B.
AU - Sykes, H.
AU - Cassady, J. R.
AU - Roemer, R. B.
N1 - Funding Information:
This research was supported by NIH grant No. CA-33922.
PY - 1990
Y1 - 1990
N2 - In this study a scanned focused ultrasound (SFUS) system was used to heat 66 tumours at various anatomical locations in 52 patients. A total of 160 treatments were given. On average, temperatures were measured in 14 or 15 locations in the scanned volume. The time-averaged temperatures over the 30 min treatment period in the best treatment of each tumour were 44.0 ± 2.4°C (mean ± SD) and 39.6 ± 1.5°C at the location of the highest and lowest sensor, respectively. On average, 39% of the sensors were above 42.5°C. When only the cases that were judged to be good candidates for the hyperthermia device were analysed, 64% of the sensors reached a temperature over 42.5°C with the highest temperature achieved being 45.9±2.3°C and the lowest 40.7 ± 1.4°C. Although the system tested has many technical limitations (for example, fixed frequency, beam geometry and power during the scan cycle), the results demonstrate that therapeutic temperatures can be achieved in many tumours. Significantly better temperatures are expected when all of the theoretical potential of scanned focused ultrasound systems has been used.
AB - In this study a scanned focused ultrasound (SFUS) system was used to heat 66 tumours at various anatomical locations in 52 patients. A total of 160 treatments were given. On average, temperatures were measured in 14 or 15 locations in the scanned volume. The time-averaged temperatures over the 30 min treatment period in the best treatment of each tumour were 44.0 ± 2.4°C (mean ± SD) and 39.6 ± 1.5°C at the location of the highest and lowest sensor, respectively. On average, 39% of the sensors were above 42.5°C. When only the cases that were judged to be good candidates for the hyperthermia device were analysed, 64% of the sensors reached a temperature over 42.5°C with the highest temperature achieved being 45.9±2.3°C and the lowest 40.7 ± 1.4°C. Although the system tested has many technical limitations (for example, fixed frequency, beam geometry and power during the scan cycle), the results demonstrate that therapeutic temperatures can be achieved in many tumours. Significantly better temperatures are expected when all of the theoretical potential of scanned focused ultrasound systems has been used.
UR - http://www.scopus.com/inward/record.url?scp=0025127334&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0025127334&partnerID=8YFLogxK
U2 - 10.3109/02656739009140971
DO - 10.3109/02656739009140971
M3 - Article
C2 - 2250115
AN - SCOPUS:0025127334
SN - 0265-6736
VL - 6
SP - 891
EP - 908
JO - International Journal of Hyperthermia
JF - International Journal of Hyperthermia
IS - 5
ER -