Abstract

“Asleep” deep brain stimulation using general anesthesia and intraoperative MR imaging guidance is considered “off-label” use by current FDA guidelines but is widely used in neurosurgical practice, and excellent safety has been demonstrated using first-generation, omnidirectional electrodes. Safety data for second-generation, directional electrodes in the interventional MR imaging environment have not yet been published. Herein, we report 34 cases of asleep deep brain stimulation using second-generation, directional electrodes in an interventional MR imaging suite at a single institution. Procedural complications and imaging data are described. All patients underwent postoperative MR imaging with fully implanted (“internalized”) electrodes after scalp closure; 4 patients also underwent MR imaging with “externalized” electrodes before scalp closure. No MR imaging-related complications were observed, and procedural complication rates were comparable to prior series. This suggests that the use of second-generation, directional electrodes in the interventional MR imaging environment appears to be safe when following manufacturer-published imaging guidelines.

Original languageEnglish (US)
Pages (from-to)2257-2262
Number of pages6
JournalAmerican Journal of Neuroradiology
Volume41
Issue number12
DOIs
StatePublished - Dec 1 2020

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Clinical Neurology

Fingerprint

Dive into the research topics of 'Safety profile of infinity deep brain stimulation electrode placement in a 1.5T interventional MRI suite: Consecutive single-institution case series'. Together they form a unique fingerprint.

Cite this