Abstract
Three critical improvements resulting from computer simulations for present day and future retinal vision implants are proposed and discussed: (1) A time profile for the stimulation current that leads predominantly to transversal stimulation of nerve cells; (2) auxiliary electric currents for electric field shaping with a time profile chosen such that these currents have small probability to cause stimulation; and (3) a local area scanning procedure that results in high pixel density for image/percept formation, except for losses at the boundary of an electrode array.
| Original language | English (US) |
|---|---|
| Article number | 7548104 |
| Pages (from-to) | 31-44 |
| Number of pages | 14 |
| Journal | IEEE Circuits and Systems Magazine |
| Volume | 16 |
| Issue number | 3 |
| DOIs | |
| State | Published - Jul 1 2016 |
| Externally published | Yes |
ASJC Scopus subject areas
- Computer Science Applications
- Electrical and Electronic Engineering