Abstract
Purpose: To investigate using 3D γ analysis for IMRT and VMAT QA. Methods: We explored and studied 3D γ-analysis by comparing TPS computed and EPID back-projection reconstructed doses in patient's CT images. Two 3D γ quantities, γPTV and γ10, were proposed and studied for evaluating the QA results, and compared to 2D γ (MapCheck composite: γMC). Results: It was found that when 3(global)3 mm criteria was used, all IMRT and 90 of VMAT plans passed QA with a γ pass rate 90. A significant statistical correlation was observed between 3D and 2D γ-analysis results for IMRT QA if γ10 and γMC are concerned, but no significant relation is found between γPTV and γMC. Conclusions: 3D γ analysis based on EPID dose back-projection may provide a feasible tool for IMRT and VMAT pretreatment plan QA.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 3051-3059 |
| Number of pages | 9 |
| Journal | Medical physics |
| Volume | 39 |
| Issue number | 6 |
| DOIs | |
| State | Published - Jun 2012 |
Keywords
- 3D-Gamma
- EPID
- IMRT
- QA
- VMAT
ASJC Scopus subject areas
- Biophysics
- Radiology Nuclear Medicine and imaging
Fingerprint
Dive into the research topics of 'On using 3D γ-analysis for IMRT and VMAT pretreatment plan QA'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS