Skip to main navigation Skip to search Skip to main content

On using 3D γ-analysis for IMRT and VMAT pretreatment plan QA

  • Chuan Wu
  • , Kelly E. Hosier
  • , Kristen E. Beck
  • , Marlyn B. Radevic
  • , Joerg Lehmann
  • , Hao H. Zhang
  • , Aaron Kroner
  • , Sharon C. Dutton
  • , Seth A. Rosenthal
  • , Jonathan K. Bareng
  • , Mark D. Logsdon
  • , David R. Asche

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: To investigate using 3D γ analysis for IMRT and VMAT QA. Methods: We explored and studied 3D γ-analysis by comparing TPS computed and EPID back-projection reconstructed doses in patient's CT images. Two 3D γ quantities, γPTV and γ10, were proposed and studied for evaluating the QA results, and compared to 2D γ (MapCheck composite: γMC). Results: It was found that when 3(global)3 mm criteria was used, all IMRT and 90 of VMAT plans passed QA with a γ pass rate 90. A significant statistical correlation was observed between 3D and 2D γ-analysis results for IMRT QA if γ10 and γMC are concerned, but no significant relation is found between γPTV and γMC. Conclusions: 3D γ analysis based on EPID dose back-projection may provide a feasible tool for IMRT and VMAT pretreatment plan QA.

Original languageEnglish (US)
Pages (from-to)3051-3059
Number of pages9
JournalMedical physics
Volume39
Issue number6
DOIs
StatePublished - Jun 2012

Keywords

  • 3D-Gamma
  • EPID
  • IMRT
  • QA
  • VMAT

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'On using 3D γ-analysis for IMRT and VMAT pretreatment plan QA'. Together they form a unique fingerprint.

Cite this