In silico fatigue optimization of TAVR stent designs with physiological motion in a beating heart model

Kyle Baylous, Ryan Helbock, Brandon Kovarovic, Salwa Anam, Marvin Slepian, Danny Bluestein

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

Background and Objective: The rapid expansion of TAVR to younger, low-risk patients raises concerns regarding device durability. Necessarily, extended stent lifetime will become more critical for new generation devices. In vitro methods commonly used for TAVR stent fatigue testing exclude the effects of the beating heart. We present a more realistic in silico stent fatigue analysis utilizing a beating heart model in which TAVR stents experience complex, nonuniform dynamic loading. Methods: Virtual TAVR deployments were simulated in the SIMULIA Living Heart Human Model of a beating heart using stent models of the self-expandable nitinol 26-mm CoreValve and Evolut R devices, and a 27-mm PolyV-2. Stent deformation was monitored over three cardiac cycles, and fatigue resistance was evaluated for the nitinol stents using finite element analysis via ABAQUS/Explicit. The average strain and strain amplitude of each stent element were tracked, and established thresholds were applied to determine potential fatigue failure. Fatigue performance of control stents was compared to parametrically modified models with a 20% increase or decrease in strut width. Results: Stents with reduced strut width applied lower radial force against the contracting myocardium of the beating heart, resulting in larger displacements and higher strain values. Formulas relating in vivo strain to stent design do not account for this. In all models, there were elements in which strains exceeded fatigue failure. The PolyV-2 stent had far fewer failing elements since its struts were optimized to reduce the strain in stent joints, achieving better fatigue resistance in the stent crown and waist elements. Different stent sections showed markedly different fatigue resistance due to the varying loading conditions. Conclusions: Our analysis indicates that previous studies underestimate strain amplitudes that may cause stent failure. This study demonstrates the utility of advanced in silico analysis of devices deployed within a beating heart that mimics in vivo loading, offering a cost-effective alternative to human or animal trials and establishing a platform to assess the impact of device design on device durability. The limited fatigue life of TAVR stents indicated here highlights a clinical complication that may eventually develop as younger, lower-risk TAVR patients, age.

Original languageEnglish (US)
Article number107886
JournalComputer Methods and Programs in Biomedicine
Volume243
DOIs
StatePublished - Jan 2024

Keywords

  • Beating heart
  • Durability
  • Fatigue
  • Stent
  • Strain
  • TAVR

ASJC Scopus subject areas

  • Software
  • Computer Science Applications
  • Health Informatics

Fingerprint

Dive into the research topics of 'In silico fatigue optimization of TAVR stent designs with physiological motion in a beating heart model'. Together they form a unique fingerprint.

Cite this