Skip to main navigation Skip to search Skip to main content

Evaluation of plain-film digitizers: A comparison of different systems and models

  • H. Roehrig
  • , W. J. Dallas
  • , H. Blume
  • , E. A. Krupinski
  • , M. M. Rogulski
  • , J. Fan
  • , K. Gandhi

Research output: Contribution to journalArticlepeer-review

Abstract

This paper describes the results of the comparison of 5 film digitizers designed for medical application to find an appropriate film digitizer for a project involving display of mammograms on a high performance CRT. Two of the digitizers were of the laser-scanning type while 3 were based on CCD technology. The comparison included estimates of linearity, uniformity, signal-to-noise ratio and spatial resolution in terms of the Modulation Transfer Function (MTF). The digitizers investigated show vastly different results. No system excelled in all performance parameters. A laser scanning digitizer was chosen for the planned display project because (1) it had the best linearity (down to OD 4); (2) it had the highest SNR at OD 4; (3) it had a comparatively good MTF.

Original languageEnglish (US)
Pages (from-to)694-703
Number of pages10
JournalProceedings of SPIE - The International Society for Optical Engineering
Volume4682
DOIs
StatePublished - 2002

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Instrumentation
  • Condensed Matter Physics
  • Computer Science Applications
  • Applied Mathematics
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Evaluation of plain-film digitizers: A comparison of different systems and models'. Together they form a unique fingerprint.

Cite this