Abstract
A micro-Faraday based focal plane array detector was developed for mass spectrometric applications. Advanced multiplexer electronics for direct charge detection and versatile signal integration and readout were used for the development of the detector. The Faraday array was fabricated using conventional lithography technique. The detector was assembled with a Mattauch-Herzog geometry MS (MHMS) system and tested by operating them in both static and scanning modes. The results show that the resolution obtained in the static mode was about R∼150, which was improved by scanning to R∼300.
Original language | English (US) |
---|---|
Pages | 879-880 |
Number of pages | 2 |
State | Published - 2002 |
Event | Proceedings - 50th ASMS Conference on Mass Spectrometry and Allied Topics - Orlando, FL, United States Duration: Jun 2 2002 → Jun 6 2002 |
Other
Other | Proceedings - 50th ASMS Conference on Mass Spectrometry and Allied Topics |
---|---|
Country/Territory | United States |
City | Orlando, FL |
Period | 6/2/02 → 6/6/02 |
ASJC Scopus subject areas
- Spectroscopy