Abstract
Purpose: To evaluate the robustness of TG119-based quality assurance metrics for an IMRT system. Methods: Four planners constructed treatment plans for the five IMRT test cases described in TG119. All plans were delivered to a 30 cm×30 cm×15 cm solid water phantom in one treatment session in order to minimize session-dependent variation from phantom setup, film quality, machine performance, etc. Composite measurements utilized film and an ionization chamber. Per-field measurements were collected using a diode array device at an effective depth of 5 cm. All data collected were analyzed using the TG119 specifications to determine the confidence limit values for each planner separately and then compared. Results: The mean variance of ion chamber measurements for each planner was within 1.7% of the planned dose. The resulting confidence limits were 3.13%, 1.98%, 3.65%, and 4.39%. Confidence limit values determined by composite film analysis were 8.06%, 13.4%, 9.30%, and 16.5%. Confidence limits from per-field measurements were 1.55%, 0.00%, 0.00%, and 2.89%. Conclusions: For a single IMRT system, the accuracy assessment provided by TG119-based quality assurance metrics showed significant variations in the confidence limits between planners across all composite and per-field evaluations. This observed variation is likely due to the different levels of modulation between each planner's set of plans. Performing the TG119 evaluation using plans produced by a single planner may not provide an adequate estimation of IMRT system accuracy.
Original language | English (US) |
---|---|
Pages (from-to) | 1641-1648 |
Number of pages | 8 |
Journal | Medical physics |
Volume | 38 |
Issue number | 3 |
DOIs | |
State | Published - Mar 2011 |
Keywords
- IMRT
- commissioning
- quality assurance
ASJC Scopus subject areas
- Biophysics
- Radiology Nuclear Medicine and imaging