Abstract
We have developed a dual-modality CT/SPECT imaging system for small-animal imaging applications. The X-ray system comprises a commercially available micro-focus X-ray tube and a CCD-based X-ray camera. X-ray transmission measurements are performed based on cone-beam geometry. Individual projections are acquired by rotating the animal about a vertical axis in front of the CCD detector. A high-resolution CT image is obtained after reconstruction using an ordered subsets-expectation maximization (OS-EM) reconstruction algorithm. The SPECT system utilizes a compact semiconductor camera module previously developed in our group. The module is mounted perpendicular to the X-ray tube/CCD combination. It consists of a 64 × 64 pixellated CdZnTe detector and a parallel-hole tungsten collimator. The field of view is 1 square inch. Planar projections for SPECT reconstruction are obtained by rotating the animal in front of the detector. Gamma-ray and X-ray images are presented of phantoms and mice. Procedures for merging the anatomical and functional images are discussed.
Original language | English (US) |
---|---|
Pages (from-to) | 63-67 |
Number of pages | 5 |
Journal | IEEE Transactions on Nuclear Science |
Volume | 51 |
Issue number | 1 I |
DOIs | |
State | Published - Feb 2004 |
Keywords
- Biomedical imaging
- Cadmium zinc telluride (CdZnTe)
- Multimodality imaging
- Single-photon emission computed tomography
- Small-animal imaging
ASJC Scopus subject areas
- Nuclear and High Energy Physics
- Nuclear Energy and Engineering
- Electrical and Electronic Engineering