Characterization of a Focal Plane Camera Fitted to a Mattauch-Herzog Geometry Mass Spectrograph. 2. Use with an Inductively Coupled Plasma

James H. Barnes IV, Gregory D. Schilling, Roger Sperline, M. Bonner Denton, Erick T. Young, Charles J. Barinaga, David W. Koppenaal, Gary M. Hieftje

Research output: Contribution to journalArticlepeer-review

43 Scopus citations

Abstract

A novel charge-sensitive detector array, termed the focal plane camera (FPC), has been coupled to a Mattauch-Herzog mass spectrograph (MHMS) with an inductively coupled plasma ionization source. The FPC employs an array of gold Faraday cups, each with its own charge. integrating circuit that allows the simultaneous detection of several m/z ratios. The ion-sampling interface of the MHMS has been redesigned to provide better heat transfer away from the sampler and skimmer cones and to reduce the negative effects of turbulent gas flows around the plasma. The instrument has produced limits of detection in the tens to hundreds of parts per quadrillion regime and isotope ratio accuracy and precision of 5% error and 0.007% RSD, respectively. Limits of detection with the FPC are comparable to those obtained with a single-channel secondary electron multiplier (SEM). However, the isotope ratio accuracy and precision are better with the FPC than when the SEM is employed. The dynamic range has been shown to be linear over 7 orders of magnitude.

Original languageEnglish (US)
Pages (from-to)2531-2536
Number of pages6
JournalAnalytical Chemistry
Volume76
Issue number9
DOIs
StatePublished - May 1 2004
Externally publishedYes

ASJC Scopus subject areas

  • Analytical Chemistry

Fingerprint

Dive into the research topics of 'Characterization of a Focal Plane Camera Fitted to a Mattauch-Herzog Geometry Mass Spectrograph. 2. Use with an Inductively Coupled Plasma'. Together they form a unique fingerprint.

Cite this