TY - JOUR
T1 - Assessing ectasia susceptibility prior to LASIK
T2 - The role of age and residual stromal bed (RSB) in conjunction to Belin-Ambrósio deviation index (BAD-D)
AU - Ambrósio, Renato
AU - Ramos, Isaac
AU - Lopes, Bernardo
AU - Canedo, Ana Laura Caiado
AU - Correa, Rosane
AU - Guerra, Frederico
AU - Luz, Allan
AU - Price, Francis W.
AU - Price, Marianne O.
AU - Schallhor, Steven
AU - Belin, Michael W.
PY - 2014
Y1 - 2014
N2 - Purpose: To compare the ability to detect preoperative ectasia risk among LASIK candidates using classic ERSS (Ectasia Risk Score System) and Pentacam Belin-Ambrósio deviation index (BAD-D), and to test the benefit of a combined approach including BAD-D and clinical data. Methods: A retrospective nonrandomized study involved preoperative LASIK data from 23 post-LASIK ectasia cases and 266 stable-LASIK (follow up > 12 months). Preoperative clinical and Pentacam (Oculus; Wetzlar, Germany) data were obtained from all cases. Mann-Whitney's test was performed to assess differences between groups. Stepwise logistic regression was used for combining parameters. The areas under the Receiver Operating Characteristic (ROC) curves (AUC) were calculated for all parameters and combinations, with pairwise comparisons of AUC (DeLong's method). Results: Statistically significant differences were found for age, residual stromal bed (RSB), central corneal thickness and BAD-D (p<0.001), but not for sphere, cylinder or spherical equivalent (p>0.05). ERSS was 3 or more on 12/23 eyes from the ectasia group (sensitivity = 52.17%) and 48/266 eyes from the stable LASIK group (18% false positive). BAD-D had AUC of 0.931 (95% CI: 0.895 to 0.957), with cut-off of 1.29 (sensitivity = 87%; specificity = 92.1%). Formula combining BAD-D, age and RSB provided 100% sensitivity and 94% specificity, with better AUC (0.989; 95% CI: 0.969 to 0.998) than all individual parameters (p>0.001). Conclusion: BAD-D is more accurate than ERSS. Combining clinical data and BAD-D improved ectasia susceptibility screening. Further validation is necessary. Novel combined functions using other topometric and tomographic parameters should be tested to further enhance accuracy.
AB - Purpose: To compare the ability to detect preoperative ectasia risk among LASIK candidates using classic ERSS (Ectasia Risk Score System) and Pentacam Belin-Ambrósio deviation index (BAD-D), and to test the benefit of a combined approach including BAD-D and clinical data. Methods: A retrospective nonrandomized study involved preoperative LASIK data from 23 post-LASIK ectasia cases and 266 stable-LASIK (follow up > 12 months). Preoperative clinical and Pentacam (Oculus; Wetzlar, Germany) data were obtained from all cases. Mann-Whitney's test was performed to assess differences between groups. Stepwise logistic regression was used for combining parameters. The areas under the Receiver Operating Characteristic (ROC) curves (AUC) were calculated for all parameters and combinations, with pairwise comparisons of AUC (DeLong's method). Results: Statistically significant differences were found for age, residual stromal bed (RSB), central corneal thickness and BAD-D (p<0.001), but not for sphere, cylinder or spherical equivalent (p>0.05). ERSS was 3 or more on 12/23 eyes from the ectasia group (sensitivity = 52.17%) and 48/266 eyes from the stable LASIK group (18% false positive). BAD-D had AUC of 0.931 (95% CI: 0.895 to 0.957), with cut-off of 1.29 (sensitivity = 87%; specificity = 92.1%). Formula combining BAD-D, age and RSB provided 100% sensitivity and 94% specificity, with better AUC (0.989; 95% CI: 0.969 to 0.998) than all individual parameters (p>0.001). Conclusion: BAD-D is more accurate than ERSS. Combining clinical data and BAD-D improved ectasia susceptibility screening. Further validation is necessary. Novel combined functions using other topometric and tomographic parameters should be tested to further enhance accuracy.
KW - Corneal diseases/etiology
KW - Corneal topography
KW - Dilatation pathologic/etiology
KW - Keratomileusis laser in situ/adverse effects
UR - http://www.scopus.com/inward/record.url?scp=84896926032&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84896926032&partnerID=8YFLogxK
U2 - 10.5935/0034-7280.20140018
DO - 10.5935/0034-7280.20140018
M3 - Article
AN - SCOPUS:84896926032
SN - 0034-7280
VL - 73
SP - 75
EP - 80
JO - Revista Brasileira de Oftalmologia
JF - Revista Brasileira de Oftalmologia
IS - 2
ER -