Abstract
Small-animal imaging systems are often characterized using phantoms, which may not predict performance in clinical applications. An implantable synthetic SPECT lesion would facilitate characterization of lesion detectability in a living animal. Methods: Anion-exchange columns with bed volumes of 100-300 nL were constructed from medical-grade polyvinyl chloride tubing and resin. The columns were tested in an excised mouse femur and implanted in the femur of a living mouse. Imaging was performed using a prototype dual-modality SPECT/CT system. Results: Activity of 7.4-22.2 MBq (0.2-0.6 mCi) localized within the synthetic lesion. The synthetic lesions were reused multiple times. Mice tolerated the implanted columns without complications for up to 8 wk. Conclusion: A reusable, synthetic SPECT lesion was constructed and implanted in the femur of a living mouse. The synthetic lesion is useful for the development of imaging schemes and for more realistically evaluating imaging-system performance in the context of a living animal.
Original language | English (US) |
---|---|
Pages (from-to) | 1796-1799 |
Number of pages | 4 |
Journal | Journal of Nuclear Medicine |
Volume | 48 |
Issue number | 11 |
DOIs | |
State | Published - Nov 1 2007 |
Keywords
- Implantable device
- Phantom
- Small-animal SPECT
- Synthetic lesion
ASJC Scopus subject areas
- Radiology Nuclear Medicine and imaging