Abstract
The activity of free and immobilized glucose oxidase was determined using a sandwich type thin-layer electrochemical cell. The thin-layer cell consisted of a gold electrode deposited on a glass microscope slide, 165 μm thick Teflon TFE spacers, and a glass cover. Enzyme activity was determined by using cyclic voltammetry to measure the amount of hydrogen peroxide produced in the glucose oxidase catalyzed redox reaction between glucose and oxygen in the thin-layer cell. The specific activity of 13.4 nM glucose oxidase in 0.2 M aqueous sodium phosphate, pH 5.2 at room temperature, was calculated to be 4.34 U/mg GOx. Under the same experimental conditions, qualitative detection of the activity of glucose oxidase covalently immobolized to a thin radiofrequency plasma modified poly(etherurethaneurea) film was demonstrated.
Original language | English (US) |
---|---|
Pages (from-to) | 655-668 |
Number of pages | 14 |
Journal | Annals of Biomedical Engineering |
Volume | 21 |
Issue number | 6 |
DOIs | |
State | Published - Nov 1993 |
Externally published | Yes |
Keywords
- Activity
- Immobilization
- Plasma polymer
ASJC Scopus subject areas
- Biomedical Engineering