A rapid, specific high-performance liquid chromatographic analysis of theophylline in plasma, serum, and saliva is described. Proteins present in the biological samples are precipitated with 6% perchloric acid and the clear supernatant is chromatographed on a reversed-phase column. Only 100 μL of serum is required and concentrations as low as 0.07 μg/mL can be measured accurately. Other xanthines do not interfere in the assay. Within- and betweenday variation is ≤ 2.2%. The method shows less bias and greater precision than the TDx (Abbott Diagnostics) procedure commonly used in clinical laboratories.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 303-306 |
| Number of pages | 4 |
| Journal | Journal of Chromatographic Science |
| Volume | 28 |
| Issue number | 6 |
| DOIs | |
| State | Published - Jun 1990 |
ASJC Scopus subject areas
- Analytical Chemistry