A fast microflow apparatus

George H. Czerlinski, Michael Isaacson, Richard Ramires, Shu Chen, Bharath Kanagal, Chaitanya Venkatesh, Linda S. Powers

Research output: Contribution to journalArticlepeer-review

Abstract

A microflow apparatus has been built with a time resolution of 1 ms for mixing volumes of at least 80 μl per syringe. A time resolution of 3 ms is attained for mixing volumes of 10 μl per syringe. Equal volumes are mixed and concentration changes are observed photometrically using fiber optics for coupling light into and out of the observation tube having 1 mm diameter and 10 mm light path length. The reaction of ferrocytochrome c with ferrihexacyanide was used to calibrate the apparatus at 550 nm (10 nm half-width).

Original languageEnglish (US)
Pages (from-to)2604-2610
Number of pages7
JournalReview of Scientific Instruments
Volume66
Issue number3
DOIs
StatePublished - 1995

ASJC Scopus subject areas

  • Instrumentation

Fingerprint

Dive into the research topics of 'A fast microflow apparatus'. Together they form a unique fingerprint.